FIRST-PL: Commissioning the first visible photonic lantern spectrograph for sub-diffraction-limit astronomy on Subaru/SCExAO
概要
FIRST-PL (Fibered Imager foR a Single Telescope - Photonic Lantern) is a newly commissioned visible-light instrument on Subaru/SCExAO achieving spectroscopy below the diffraction limit. The instrument uses a Photonic Lantern (PL)-converting multimode fiber into 19 single-mode outputs-feeding a mid-resolution spectrograph (R 3000, 630-790 nm). On-sky performance demonstrates 40% injection efficiency at 680 nm (Strehl 30%) and 12x throughput improvement over single-mode fibers. Three operational m…